DLMD700-HA |
DLMD1000-HA |
||||||
|
Measuring range |
External Dimensions(mm):0-700 |
External Dimensions(mm):0-1000 |
|||||
|
Inner Dimensions(mm):12-580 |
Inner Dimensions(mm):12-920 |
||||||
|
Resolution |
0.05μm |
||||||
|
Indication error |
±(0.2+L/1000)μm,L is the measured length, unit:mm |
||||||
|
Repeatability |
External Dimensions:0.2μm |
||||||
|
Inner Dimensions:0.5μm |
|||||||
|
Measuring force |
(0.5-2.5)N Adjustable gear shift |
||||||
|
Universal workbench |
Vertical travel |
Horizontal travel |
Horizontal Floating |
Rotation |
Swing |
Workbench load |
|
|
0-70mm |
0-40mm |
10mm |
±3° |
±3° |
30kg |
||
|
Thread measurement |
External thread maximum:M100 |
||||||
|
Abbe axis movement range |
(0-16)mm |
||||||
|
Max. clamping diameter of the center rest |
Gauge block support (mm): Supporting gauge block length range: up to 1000 |
||||||
|
Measuring rod support (mm): supporting measuring rod length range: up to 1000 |
|||||||
|
Gauge block wringing fixture |
Wringing gauge block length range: up to 100mm |
||||||
|
Working environment requirements |
Room temperature:(20±0.4)℃,Shall not exceed 0.3℃ per hour |
||||||
|
Calibration environment requirements |
Room temperature:(20±0.2)℃,Shall not exceed 0.2℃ per hour; (Humidity, temperature, and vibration meet the measurement conditions) |
||||||
|
Power supply |
(220±22)V,No interference source, good grounding |
||||||
Optional Accessories DLMD-A
Side Hook, Elastic Measuring Hook, Universal Measuring Hook, Center Frame, Cross Bracket, Floating Worktable, Photoelectric Measuring Device, Photoelectric Measuring Hook, Standard Ring Gauge, Inclined Fixing Bracket, Gauge Block Wringing Stand, Single V-Block, Dial Gauge Stand, Three-Pin Stand, Spacer Bar